Time of Flight (TOF) - Positron Emission Tomography (PET) imaging using 5-gap Glass Multi-gap Resistive Plate Chambers (MRPCs)
Loading...
Files
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Time of Flight (TOF) technique is used in Positron Emission Tomography (PET) imaging
worldwide. PET imaging technique, is based on two back to back photons created by the
annihilation of positron (e+) (generated by FDG (radio-tracer) inside the patient's blood
stream) with electron. The scintillator based detectors currently in use are very costly, so to
minimise the cost, we are developing a different type of detector called Multi-gap Resistive
Plate Chamber (MRPCs). MRPCs are very low cost detectors and can be fabricated over
large areas having time resolution of ~10 ps. Hence they can be a suitable replacement to
scintillators for PET imaging.
In the current work, we have fabricated and characterised two identical (18 cm x 18 cm) 5-
gap glass based MRPCs. The fabricated MRPCs were kept one over the other separated by a
distance of 40cm. In our experiment, the source (Na-22) position was shifted physically by
~12 cm and the data from MRPCs nearly matched with the same. This proves the suitability
of the fabricated MRPCs for TOF measurements.