Complaints Handling and Vigilance Reporting for Stryker Neurovascular Products
Loading...
Date
Authors
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
The main goal of the company is to offer its customers the best quality products possible. One of
the best ways to monitor the quality of product and provide feedback to the design, manufacturing,
technical support, sales, and marketing teams is a well-designed Complaint
Handling System. Medical device companies can continuously enhance user safety, regulatory
compliance, and customer satisfaction by using an efficient complaint handling system. There are
comprehensive quality management system software’s available to automate functions such as
recording customer complaints, investigations, and adverse event reporting according to regulatory
standards. Trackwise is one such software which is used by Stryker for the Complaints Handling
process. Medical device companies are required by the FDA to report incidents involving medical
equipment in order to detect and address problems in a timely manner. Despite the best planning
of medical devices, sometimes adverse events occur. These events may include medical device
malfunctions, serious injuries, and deaths. Similar way events are captured by other medical device
industries and reported to FDA. The data is readily available in public database forum. The data is
extracted for market analysis and is documented in periodic safety update report (PSUR) along
with other requirements. Furthermore, this project also focuses on the establishment of an
automated statistical tool for estimating the product usage in the field.
Stryker's complaint handling system, Trackwise, serves as a central location for all complaints and
supporting data. In the complaint handling team various roles are allocated to manage complaints,
such as CIC (Complaint Intake Center), CMC (Complaint Management Center), and
PAC (Product Assessment Center). This project focuses on CMC. After the CIC intake, CMC
manages the complaint and uploads it to Trackwise. Additional event information is received via
the RAQA or Sales Representative. Files an eMDR for Reportable Complaints, submit
Complaints, approve Complaints after PAC investigation, and closes Complaints.
The Periodic Safety Update Report is a periodic report that medical device manufacturers are
required to maintain and submit to their Notified Body (NB) on a regular basis. The PSUR is a
summary of the findings of Post Market Surveillance activities, as well as the conclusions reached
by manufacturers for Class IIa, IIb and III devices. A description and justification for any
corrective or preventative actions (CAPAs) made by the manufacturer must also be included in
xviii
this report. Present study encompasses the retrieval, collection and analysis of data for periodic
safety update report of medical device Catheter.
Stryker Neurovascular does not have an established statistical method to estimate product
consumption in the market, when a product change has been made and a mixture of new and old
product are in circulation. This project focuses on establishing a tool for estimating product usage
in the field based on available data that eliminates the need for ad-hoc, manual analyses.